This report is submitted beyond 30-calendar days from allergan¿s receipt due to the exemption transition period.A review of the device history record has been completed.No deviations or non-conformances noted.Device evaluation: visual analysis of the returned device identified: fold creases, deformation and was observed after autoclave cycle: cloudy gel.A weight test of the explanted material was verified and it was within specification.Based on the device analysis the final assessment is: no observations found related with the complaint reported by the physician.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Reason for reoperation: capsular contracture, baker grade iii/iv.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.This is a known potential adverse event addressed in the product labeling.
|