|
On (b)(6) 2019, this patient underwent endovascular repair of an abdominal aortic aneurysm using gore® excluder® aaa endoprostheses.During the case a contralateral leg component (plc121400/20569329) was implanted.An iliac extender component (pll161007/1878556) was needed to complete appropriate coverage from the contralateral leg component into the left external iliac artery.Upon removal of the delivery catheter, the olive tip and approximately 3-4" of the polyimide tubing separated from the rest of the delivery catheter.The physician noted resistance as the olive tip was being pulled into the sheath (dsf1833/unk).An audible sound was heard just as the catheter popped back into the sheath.The catheter delivery system was removed.The broken portion of the catheter and the olive tip were still over the wire (boston scientific amplatz, 260cm,.035 in).The sheath was advanced over the olive tip.The sheath, wire, and catheter were pulled back simultaneously until the sheath and broken piece were outside the patient's body.Once the broken piece was determined by examination and fluoroscopy to be intact and out of the body, the case proceeded in the usual fashion.The delivery catheter and sheath will be sent to gore for evaluation.
|