• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. ULTRASONIC GASTROVIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. ULTRASONIC GASTROVIDEOSCOPE Back to Search Results
Model Number GF-UC140P-AL5
Medical Device Problem Code Microbial Contamination of Device (2303)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The subject device has not been returned to omsc but was returned to olympus (b)(4).((b)(4)).(b)(4) sent the device to a third party laboratory for microbiological testing.As a result of the testing, no microbe was detected from the sample collected from the subject device.The testing result cleared the (b)(6) guideline.The exact cause of the reported event could not be conclusively determined at this time.If additional information becomes available, this report will be supplemented.
 
Event or Problem Description
Olympus medical systems corp.(omsc) was informed that as a result of routine microbiological testing by the user facility, the sample collected from the subject device tested positive for unspecified microbes (>256 cfu / 100ml).The device had been reprocessed with a non-olympus automated endoscope reprocessor, wassenburg, using peracetic acid.There was no report of infection associated with this report.
 
Additional Manufacturer Narrative
This supplemental report is being submitted to provide additional information.Omsc reviewed the manufacturing history (dhr) of the subject device and confirmed no irregularity.The exact cause of the reported event could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRASONIC GASTROVIDEOSCOPE
Common Device Name
ULTRASONIC GASTROVIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key8868559
Report Number8010047-2019-02879
Device Sequence Number14050984
Product Code ODG
UDI-Device Identifier04953170355431
UDI-Public04953170355431
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K011314
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberGF-UC140P-AL5
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 07/18/2019
Supplement Date Received by Manufacturer09/13/2019
Initial Report FDA Received Date08/07/2019
Supplement Report FDA Received Date10/08/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No