• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARD® WOVEN URETERAL CATHETER; MEC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARD® WOVEN URETERAL CATHETER; MEC Back to Search Results
Model Number 36303
Medical Device Problem Code Inaccurate Flow Rate (1249)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
 
Event or Problem Description
It was reported that the urine sometimes backs up into the catheter.
 
Additional Manufacturer Narrative
The reported event was inconclusive as no sample was returned for evaluation.A potential failure mode could be ¿tip wall thickness too thin¿.A potential root cause for this failure could be "improper manufacturing technique/s".The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: "contraindication do not use on irritated or compromised skin.Precaution do not use if allergic reaction occurs.For good hygiene, change catheter daily.Use of a single device for longer periods than 24 hours may increase the risk of complications.1) wash penis with mild soap and warm water.Dry thoroughly.2) trim pubic hair if necessary.3) open package at perforation.4) to remove plastic insert, squeeze catheter at the top of the white cone and pull to release.5) unroll self-adhering catheter over penis.6) gently squeeze the catheter to properly seal adhesive to the skin.7) connect to collection bag.Important: wear time may be significantly reduced if adhesive is not properly sealed to the skin.Directions: to remove gently roll catheter off the penis.Note: if necessary, apply a warm wet compress (such as a wet washcloth) around the catheter to help loosen the adhesive." section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that the urine sometimes backs up into the catheter.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® WOVEN URETERAL CATHETER
Common Device Name
MEC
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key8870509
Report Number1018233-2019-04555
Device Sequence Number11294774
Product Code NNX
UDI-Device Identifier00801741071065
UDI-Public(01)00801741071065
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer,distributor,other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 08/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number36303
Device Catalogue Number36303
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/19/2019
Supplement Date Received by Manufacturer08/11/2019
Initial Report FDA Received Date08/07/2019
Supplement Report FDA Received Date08/19/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-