Catalog Number 466F220A
Device Problems Failure to Align (2522); Migration (4003)
Patient Problems Embolus (1830); Perforation of Vessels (2135)
Event Date 07/11/2019
Event Type  Injury  
Manufacturer Narrative
Occupation: other, senior counsel, litigation.Date of event: please note that the exact event date is unknown and the event date is the complaint awareness date.Description of problem or event: as reported, the patient underwent placement of an optease inferior vena cava (ivc) filter.The indication for filter placement is not available.The report states that the filter subsequently malfunctioned and caused injury and damage to the patient including, but not limited to filter migration, filter tilt, shortness of breath, and failure to stop blood clots from going through the filter.The filter remains implanted; thus, unavailable for analysis.The product was not returned for analysis and the sterile lot number has not been provided; therefore, no device analysis nor device history record review could be performed.The optease vena cava filter is indicated for use in the prevention of recurrent pulmonary embolism (pe) via percutaneous placement in the vena cava for patients in which anticoagulants are contraindicated, anticoagulant therapy for thromboembolic disease has failed, emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced or for chronic, recurrent pulmonary embolism where anticoagulant therapy has failed, or is contraindicated.The purpose of a vena cava filter is to catch thrombus from the lower extremities as it travels along normal blood flow patterns up towards the heart.Without procedural films for review, the filter tilt reported could not be confirmed.Additionally, the timing and mechanism of the filter tilt is unknown.Ivc filter tilt has been associated with the anatomy of the vessel, specifically asymmetry and tortuousness.Inferior vena cava (ivc) filter migration is a known potential adverse event associated with all ivc filter implants and is listed in the instruction for use (ifu) as such.Possible causes for filter migration include mega cava, wire entrapment during central venous catheter placement, ¿sail¿ effect (cranial migration) of large clot burden within the filter, mechanical device failure, and operator error.Physiologic causes of migration may result from temporary dysmorphism of the inferior vena cava including bending, coughing or valsalva maneuvers resulting in dislodgment of the filter.Some studies suggest that strenuous physical activity and increased intra-abdominal pressure can lead to migration of ivc filters.Blood clots within vasculature do not represent a device malfunction.Shortness of breath does not represent a device malfunction and may be related to underlying patient related issues.Clinical factors that may have influenced the event include patient, pharmacological and lesion characteristics.Without procedural films or images for review the reported event(s) could not be confirmed.Given the limited information available for review at this time, there is nothing to suggest that the reported events are related to the design and manufacturing process of the device; therefore, no corrective action will be taken.Should additional information become available, the file will be updated accordingly.Please note that this is the initial/final report for this product.
 
Event Description
As reported by the legal brief, the patient underwent placement of a optease vena cava filter.The report states that the filter subsequently malfunctioned and caused injury and damage to the patient including, but not limited to: filter migration, filter tilt, shortness of breath, and failure to stop blood clots from going through the filter.
 
Manufacturer Narrative
As reported, the patient underwent placement of an optease retrievable vena cava filter.The patient is reported to have had a history of pulmonary embolism (pe) and left lower extremity deep vein thrombosis (dvt) that extended from the distal superficial femoral vein to the common femoral vein.The filter was implanted via the right common femoral artery and placed distal to the renal veins without complications.This was followed by thrombectomy and thrombolysis of a dvt in the left lower extremity.At some point after the filter implantation, the patient became aware that the filter had tilted and migrated.The filter had failed to stop blood clots from going through the filter and was associated with perforation of filter struts outside the inferior vena cava (ivc).In addition, the patient reported having undergone an unsuccessful retrieval attempt; though a retrieval attempt was not documented.The patient further reported having experienced shortness of breath, continuous discomfort and mental anguish associated with the filter.The product was not returned for analysis.A review of the device history record revealed no anomalies during the manufacturing and inspection processes that can be associated with the reported complaint.The optease retrievable vena cava filter is indicated for use in the prevention of recurrent pe via percutaneous placement in the ivc for patients in which anticoagulants are contraindicated, anticoagulant therapy for thromboembolic disease has failed, emergency treatment following massive pe where anticipated benefits of conventional therapy are reduced or for chronic, recurrent pe where anticoagulant therapy has failed, or is contraindicated.The purpose of a vena cava filter is to catch thrombus from the lower extremities as it travels along normal blood flow patterns up towards the heart.Without images or procedural films for review, the reported filter tilt, migration and ivc perforation events could not be confirmed and the exact cause could not be determined.Ivc filter tilt has been associated with the anatomy of the vessel, specifically asymmetry and tortuousness.Additionally, the timing and mechanism of the filter tilt is unknown.Ivc filter migration is a known potential adverse event associated with all ivc filter implants and is listed in the instruction for use (ifu) as such.Possible causes for filter migration include mega cava, wire entrapment during central venous catheter placement, ¿sail¿ effect (cranial migration) of large clot burden within the filter, mechanical device failure, and operator error.Physiologic causes of migration may result from temporary dysmorphism of the ivc including bending, coughing or valsalva maneuvers resulting in dislodgment of the filter.Some studies suggest that strenuous physical activity and increased intra-abdominal pressure can lead to migration of ivc filters.It is unknown if the tilt contributed to the reported perforation.A review of the ifu notes vessel damage such as intimal tears and perforation as procedural complications related to ivc filters.Perforation from removable filters is relatively common, and directly related to how long the filter has been in place.Studies have noted a greater than 80% perforation rate overall, with all filters imaged after 71 days from implantation revealing some level of perforation.Blood clots and thrombosis and/or occlusion within the device or within the ivc and/or vasculature do not represent a device malfunction.Due to the nature of the complaint, the reported shortness of breath and pain experienced by the patient could not be confirmed and the exact cause could not be determined.These clinical events do not represent evidence of a device malfunction.Clinical factors that may have influenced these events include the patient¿s pre-existing co-morbidities, pharmacological issues and lesion characteristics.The anxiety experienced by the patient does not represent a device malfunction.Anxiety, part of the body¿s natural response to stress and can cause feelings of, but not limited to, nervousness, mental anguish, fear, unease and worry.Given the limited information available for review, there is nothing to suggest that a malfunction in the design and manufacturing process of the device; therefore, no corrective action will be taken.Should additional information become available, the file will be updated accordingly.
 
Event Description
As reported by the legal brief, the patient underwent placement of a optease vena cava filter.The report states that the filter subsequently malfunctioned and caused injury and damage to the patient including, but not limited to: filter migration, filter tilt, shortness of breath, and failure to stop blood clots from going through the filter.Additional information received per the patient profile form (ppf) states that the patient experienced perforation of filter struts outside the inferior vena cava (ivc) and tilting of the filter.The patient also experienced continuous discomfort and mental anguish.The form states there was an unsuccessful removal which left the device in place; however, the form also states that there has been no attempt to remove the device.Per the patient¿s implant records, the patient had a history of pulmonary embolism with deep vein thrombosis (dvt) and needed a dvt thrombectomy.The patient already had pulmonary embolism, and prior to the dvt thrombectomy, the ivc filter was placed.The optease filter was deployed distal to the renal veins under fluoroscopic guidance.There were no complications.Upon filter placement, a deep venous thrombectomy of the left lower extremity was completed.
 
Manufacturer Narrative
After further review of additional information received, the following sections have been updated accordingly: b4, d1, d4, g4, g7, h1, h2, h4 and h6.Section b5: additional information received per the patient profile form (ppf) states that the patient experienced perforation of filter struts outside the inferior vena cava (ivc) and tilting of the filter.The patient also experienced continuous discomfort and mental anguish.The form states there was an unsuccessful removal attempt which left the device in place; however, the form also states that there has been no attempt to remove the device.As reported, the patient underwent placement of an optease inferior vena cava (ivc) filter.The indication for filter placement is not available.The report states that the filter subsequently malfunctioned and caused injury and damage to the patient including, but not limited to filter migration, tilt, shortness of breath, and failure to stop blood clots from going through the filter.Per the patient profile form (ppf), the patient experienced perforation of filter struts outside the inferior vena cava (ivc) and tilting of the filter.The patient also reports continuous discomfort and mental anguish.The form reports both an unsuccessful removal and that there has been no attempt to remove the device.The product was not returned for analysis.A review of the device history record revealed no anomalies during the manufacturing and inspection processes that can be associated with the reported complaint.The optease vena cava filter is indicated for use in the prevention of recurrent pulmonary embolism (pe) via percutaneous placement in the vena cava for patients in which anticoagulants are contraindicated, anticoagulant therapy for thromboembolic disease has failed, emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced or for chronic, recurrent pulmonary embolism where anticoagulant therapy has failed, or is contraindicated.The purpose of a vena cava filter is to catch thrombus from the lower extremities as it travels along normal blood flow patterns up towards the heart.Without procedural films for review, the filter tilt reported could not be confirmed.Additionally, the timing and mechanism of the filter tilt is unknown.Ivc filter tilt has been associated with the anatomy of the vessel, specifically asymmetry and tortuousness.It was reported that there was perforation of the ivc; however, a clinical conclusion could not be determined as to the cause of the event.A review of the instructions for use notes vessel damage such as intimal tears and perforation as procedural complications related it ivc filters.Ivc perforation from removable filters is relatively common, and directly related to how long the filter has been in place.Studies have noted a greater than 80% perforation rate overall, with all filters imaged after 71 days from implantation revealing some level of perforation.Inferior vena cava (ivc) filter migration is a known potential adverse event associated with all ivc filter implants and is listed in the instruction for use (ifu) as such.Possible causes for filter migration include mega cava, wire entrapment during central venous catheter placement, ¿sail¿ effect (cranial migration) of large clot burden within the filter, mechanical device failure, and operator error.Physiologic causes of migration may result from temporary dysmorphism of the inferior vena cava including bending, coughing or valsalva maneuvers resulting in dislodgment of the filter.Some studies suggest that strenuous physical activity and increased intra-abdominal pressure can lead to migration of ivc filters.Blood clots and occlusive thrombosis within the filter and vasculature do not represent a device malfunction.Anxiety, shortness of breath and pain do not represent a device malfunction and may be related to underlying patient related issues.Clinical factors that may have influenced the event include patient, pharmacological and lesion characteristics.Without procedural films or images for review the reported event(s) could not be confirmed.Given the limited information available for review at this time, there is nothing to suggest that the reported events are related to the design and manufacturing process of the device; therefore, no corrective action will be taken.Should additional information become available, the file will be updated accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTEASE RETRIEVAL FILTER
Type of Device
FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
CORDIS CASHEL
cahir road
cashel co. tipperary
EI 
MDR Report Key8871597
MDR Text Key153971691
Report Number1016427-2019-03139
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,other
Type of Report Initial,Followup,Followup
Report Date 10/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2016
Device Catalogue Number466F220A
Device Lot Number15989715
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 07/11/2019
Initial Date FDA Received08/07/2019
Supplement Dates Manufacturer Received09/19/2019
10/09/2019
Supplement Dates FDA Received10/16/2019
10/21/2019
Patient Sequence Number1
Treatment
UNK 6F SHEATH
Patient Outcome(s) Life Threatening;
Patient Age55 YR
-
-