• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE Back to Search Results
Model Number V60
Device Problem Failure to Zero (1683)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2018.Date of report: 07aug2019.The manufacturer's field service engineer (fse) performed remote troubleshooting and recommended the customer swap solenoids 2&4 noting if error changes to 1108 or 1109 to replace solenoid 4.The customer replaced the data acquisition (da) pcba to address the issue.The data acquisition (da) pcba was tested and no failures were identified.This mdr has been reassessed as reportable after a request from the fda on march 1, 2019 to review complaints from 29 nov 2017 to 29 nov 2018.As this has been reassessed, it will appear to be a late mdr.
 
Event Description
It was reported that the ventilator experienced proximal pressure sensor autozero fail.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V60 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS
3000 minuteman road
andover MA 01810
Manufacturer Contact
melissa abbott
2271 cosmos court
carlsbad, CA 92011
9093746996
MDR Report Key8872272
MDR Text Key153709461
Report Number2031642-2019-05600
Device Sequence Number1
Product Code MNT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 04/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/07/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/14/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/09/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/19/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-