• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MMJ SA DE CV(USD) SHILEY; TUBE, TRACHEAL (W/WO CONNECTOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MMJ SA DE CV(USD) SHILEY; TUBE, TRACHEAL (W/WO CONNECTOR) Back to Search Results
Model Number 86111-
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 07/22/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, the device was tested prior to use by injecting air.It was noted that there was increased resistance during both inflation and deflation.The patient was successfully intubated with another device of undetermined lot.
 
Additional Manufacturer Narrative
Additional info: d10, g4, h3, h6 h3 evaluation summary: two samples of endotracheal tube were received for evaluation.An inflation/deflation test was performed on both samples 25cc of air was applied to both cuffs using a syringe and it was observed inflation was difficult and deflation was hard on both tubes, it was noticed the air has difficulties to get through the inflation systems in the area where the inflation line is inserted into the tubes, the failure mode cuff won¿t inflate is confirmed.Corrective actions have been implemented to mitigate this issue.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHILEY
Common Device Name
TUBE, TRACHEAL (W/WO CONNECTOR)
Manufacturer (Section D)
MMJ SA DE CV(USD)
ave henequen no 1181 desarrol
ciudad juarez,mx 32590
MX  32590
MDR Report Key8872473
Report Number2936999-2019-00524
Device Sequence Number11282399
Product Code BTR
UDI-Device Identifier40884522000117
UDI-Public40884522000117
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K871204
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/22/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/14/2024
Device Model Number86111-
Device Catalogue Number86111-
Device Lot Number19B0312JZX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/12/2019
Initial Date Received by Manufacturer 07/22/2019
Supplement Date Received by Manufacturer09/30/2019
Initial Report FDA Received Date08/07/2019
Supplement Report FDA Received Date10/22/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-