• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND CENTRIFUGAL PUMP SYSTEM WITH TUBING CLAMP; CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND CENTRIFUGAL PUMP SYSTEM WITH TUBING CLAMP; CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 60-02-15
Medical Device Problem Code Partial Blockage (1065)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/10/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Patient information was not provided.Livanova (b)(4) manufactures the centrifugal pump system with tubing clamp.The incident occurred in (b)(6).The affected part was requested to be returned to livanova (b)(4) for a detailed investigation.Corrective actions are in progress for this issue.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event or Problem Description
Livanova (b)(4) received a report that the shaft angle encoder of a centrifugal pump system with tubing clamp intermittently locked and did not turn during procedure.There was no report of patient injury.
 
Additional Manufacturer Narrative
Additional narrative: deviation could be reproduced during the investigation.The shaft encoder was stuck.
 
Event or Problem Description
See initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRIFUGAL PUMP SYSTEM WITH TUBING CLAMP
Common Device Name
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich
MDR Report Key8872666
Report Number9611109-2019-00607
Device Sequence Number2754759
Product Code DWA
UDI-Device Identifier04033817900986
UDI-Public(01)04033817900986(11)181009
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K032213
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 09/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number60-02-15
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 07/10/2019
Supplement Date Received by Manufacturer08/21/2019
Initial Report FDA Received Date08/08/2019
Supplement Report FDA Received Date09/19/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-