| Model Number |
60-02-15 |
| Medical Device Problem Code |
Partial Blockage (1065)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/10/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Patient information was not provided.Livanova (b)(4) manufactures the centrifugal pump system with tubing clamp.The incident occurred in (b)(6).The affected part was requested to be returned to livanova (b)(4) for a detailed investigation.Corrective actions are in progress for this issue.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
| |
|
Event or Problem Description
|
|
Livanova (b)(4) received a report that the shaft angle encoder of a centrifugal pump system with tubing clamp intermittently locked and did not turn during procedure.There was no report of patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional narrative: deviation could be reproduced during the investigation.The shaft encoder was stuck.
|
| |
|
Event or Problem Description
|
|
See initial report.
|
| |
|