| Model Number |
MMT-722NAS |
| Medical Device Problem Code |
Failure to Cycle (1142)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/29/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the insulin pump alarmed compromised force sensor system.The customer¿s blood glucose level was 130 mg/dl at the time of the incident.Customer reported that they was able to clear the alarm.Customer was able to complete rewound.The customer was advised that the insulin pump needed to be replaced and to revert to the backup plan as per health care professional instruction.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device passed the displacement test but a33 alarm during the basic occlusion test due to loose/protruded drive support disk.Unable to perform the occlusion, prime/a33 and excessive no delivery test due to a33 alarm.
|
| |
|