TERUMO CARDIOVASCULAR SYSTEMS CORPORATION SARNS 8000 UNIVERSAL ROLLER PUMP; PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE ¿ SARNS 8000
|
Back to Search Results |
|
Model Number 16402 |
Device Problem
Output Problem (3005)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 07/17/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
The field service representative (fsr) verified the issue.When the device was turned on an overspeed message immediately appeared on the display and the pump would not rotate.The fsr spun the head by hand and the pump rotated making a grinding noise.The pump was removed and was returned to the manufacturer for further evaluation.
|
|
Event Description
|
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, the roller pump had an "overspeed 1" error.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
|
|
Manufacturer Narrative
|
The reported complaint was confirmed.During laboratory analysis, the product surveillance technician (pst) observed "overspeed 1" error.The motor was replaced with a lab use only motor and operated normally determining that the customer motor was the cause of the error message.The product will be sent to service to be brought to manufacturer¿s specifications before being returned to the customer.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|