• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTERVASCULAR SAS STANDARD; PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTERVASCULAR SAS STANDARD; PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER Back to Search Results
Model Number INTERGARD WOVEN AORTIC ARCH
Medical Device Problem Code Manufacturing, Packaging or Shipping Problem (2975)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(3331/213) a review of the complaint device history records, indicated that the graft was processed and inspected according to established procedures and was therefore released following acceptable quality inspections and tests.These tests include a 100% inner diameter measurement based on the international iso (b)(4) standard for vascular grafts.(4109/213) the review of post-marketing historical data indicated that no other similar complaint was reported for the same lot number.(10/213) a fragment of the involved product was returned to intervascular and was inspected by the quality assurance (qa) manager.His observations are as follows: "the sample was packaged in a pill box.The measurement of the sample diameter of the product body was performed with the qa measurement set no.1 mes 12-078 the result was: 26.8mm for a specification of 26mm +/- 1mm.From a diameter perspective, the transmitted sample conforms to the specification." (67) the conducted investigation suggests that the product was not defective.
 
Event or Problem Description
According to the surgeon, the diameter of the prosthesis is not the one indicated on the packaging.It was later confirmed that the graft was implanted.The surgeon cut a piece of the graft and sent it to us for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STANDARD
Common Device Name
PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER
Manufacturer (Section D)
INTERVASCULAR SAS
zone industrielle athelia i
la ciotat
MDR Report Key8873275
Report Number1640201-2019-00037
Device Sequence Number16791164
Product Code DSY
UDI-Device Identifier00384401010746
UDI-Public00384401010746
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K013651
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Expiration Date10/31/2023
Device Model NumberINTERGARD WOVEN AORTIC ARCH
Device Catalogue NumberHEWAA2610080810/1
Device Lot Number18L01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/02/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/15/2019
Initial Report FDA Received Date08/08/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-