| Model Number |
60-02-15 |
| Medical Device Problem Code |
Partial Blockage (1065)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
07/11/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4), united states.There was no patient involvement.Livanova (b)(4) manufactures the centrifugal pump system with tubing clamp.The incident occurred in (b)(4), united states.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.The affected parts have been asked to livanova (b)(4) for a detailed investigation.Corrective actions are in progress for this issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
| |
|
Event or Problem Description
|
|
Livanova (b)(4) received a report that the speed of a centrifugal pump system with tubing clamp could not be adjusted any more during in-service.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
Deviation could be reproduced during the investigation at the manufacturer site.The shaft encoder was stuck.
|
| |
|
Event or Problem Description
|
|
See initial report.
|
| |
|