• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHROCARE CORP. SUPER TURBOVAC 90 IFS; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHROCARE CORP. SUPER TURBOVAC 90 IFS; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Catalog Number ASH4250-01
Medical Device Problem Codes Flaked (1246); Dull, Blunt (2407)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 04/01/2019
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during right knee scope and lysis of adhesions, the wand shaft was peeling off inside the joint.Black pieces of the wand were floating in the joint therefore surgeon suctioned this out with arthroscopic shaver.The procedure was successfully completed without delay using the same device.No patient injury or other complications were reported.
 
Additional Manufacturer Narrative
The device, used in treatment, was not returned for evaluation.A relationship, if any, between the subject device and the reported event could not be determined since the product was not returned for evaluation.Visual inspection and functional testing could not be performed because the device in question was not returned for evaluation.Thus, the complaint could not be verified, nor could a root cause be determined with confidence.If the product associated with this event is returned at a future date, this evaluation will be reopened for investigation.Factors not associated with the manufacture or design of the device which could result in damage to the black shrink tube is improper insertion or removal from an apparatus, using the wand as a lever to enlarge the surgical site, or rubbing against a hard object such as bone and are outlined in the precautionary statements in the instruction for use.There were no indications during manufacturing record review that would suggest that the device did not meet product specifications upon release into distribution.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUPER TURBOVAC 90 IFS
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
MDR Report Key8873474
Report Number3006524618-2019-00375
Device Sequence Number2390056
Product Code GEI
UDI-Device Identifier00817470006506
UDI-Public00817470006506
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K083306
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/10/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberASH4250-01
Device Lot Number2026542
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/15/2019
Supplement Date Received by Manufacturer09/09/2019
Initial Report FDA Received Date08/08/2019
Supplement Report FDA Received Date09/10/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age65 YR
-
-