| Catalog Number |
368661 |
| Medical Device Problem Codes |
Leak/Splash (1354); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/03/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr: a device evaluation and/or device history review is anticipated, but is not complete.Upon completion, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It has been reported that the bd vacutainer® k2 edta (k2e) 10.8 mg blood collection tubes has been found experiencing molding defects.The following has been provided by the initial reporter: it was reported purple top tubes were deformed.This is to inform you on (b)(6) 2019 staff discovered that about 15-20 edta 6ml purple top tubes were deformed, from lot number 8156830.Verbatim: during qc function checks, it was discovered that about 15-20 edta 6ml purple top tubes were deformed, from lot number 8156830.
|
| |
|
Event or Problem Description
|
|
It has been reported that the bd vacutainer® k2 edta (k2e) 10.8 mg blood collection tubes has been found experiencing molding defects.The following has been provided by the initial reporter: it was reported purple top tubes were deformed.This is to inform you on (b)(6) 2019 staff discovered that about 15-20 edta 6ml purple top tubes were deformed, from lot number 8156830.Verbatim in attachment: during qc function checks, it was discovered that about 15-20 edta 6ml purple top tubes were deformed, from lot number 8156830.
|
| |
|
Additional Manufacturer Narrative
|
|
H.6.Investigation: bd had not received samples or photos from the customer facility for evaluation; therefore, the investigation was limited.A review of the device history record was completed for the incident lot and, based on this review, all product specifications and requirements for lot release were met; there were no related quality non-conformances during manufacturing of the product.Note that there are potentially varying degrees of tube collapse, based on the temperature and duration of time at elevated temperature.Mildly collapsed tubes will retain vacuum and only show minimal deformation.Mildly collapsed tubes will draw appropriately, but may not be able to be processed on laboratory instruments (e.G.May be slightly bowed and no longer fit in a rack).As long as the vacuum is retained and the tube fills appropriately, the tube will function properly.Fully collapsed tubes appear flattened, twisted, and visibly deformed.Fully collapsed tubes retain little or no vacuum and therefore cannot be drawn to an appreciable volume.These tubes are easily identified before use by the healthcare worker, so there is no potential impact to the patient.H3 other text : see h.10.
|
| |
|