• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION PROCLAIM SCS IPG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION PROCLAIM SCS IPG Back to Search Results
Model Number 3665
Medical Device Problem Code Application Program Freezes or Fails to Launch (4031)
Health Effect - Clinical Code Inadequate Pain Relief (2388)
Date of Event 07/18/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device is included in the neuromodulation implantable pulse generator (ipg) inoperable when exposed to monopolar electrosurgery advisory notice issued by abbott on 02 june 2017.
 
Event or Problem Description
It was reported the patient underwent a surgical procedure for a wound cleaning in her foot.Reportedly, after the procedure the scs system patient controller (pc) no longer communicated with the scs ipg.The system had only been switched off, but not set to surgery mode.A company representative attempted to connect with clinical programmer but a message: "unable to connect to the generator.A problem has been detected." was displayed.Consequently, surgical intervention may be necessary to replace the patient's scs ipg.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROCLAIM SCS IPG
Common Device Name
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
MDR Report Key8876204
Report Number1627487-2019-08971
Device Sequence Number2587044
Product Code LGW
UDI-Device Identifier05415067023247
UDI-Public05415067023247
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
P010032
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Physician
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 09/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/12/2020
Device Model Number3665
Device Lot Number6401293
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 07/18/2019
Supplement Date Received by Manufacturer09/11/2019
Initial Report FDA Received Date08/08/2019
Supplement Report FDA Received Date09/11/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
1627487/06/02/2017/001-C
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-