Model Number SA-85
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code Injury (2348)
Date of Event 04/16/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
A heli-fx endoanchoring system was implanted for the treatment of neck dilation in an endurant stent graft system in a 79.2 abdominal aortic aneurysm.An endurant ii limb extension was also implanted during the procedure.The endurant stent graft system was originally implanted 5 years earlier.The proximal aortic neck measured 27.8mm to 33.9mm.The neck length was noted as 7.5mm.It was reported that 2.5 years post endoanchor implantation a gutter endoleak from the left renal artery stent was observed.Intervention with balloon angioplasty of proximal endograft seal zone and kissing balloon angioplasty of the left renal artery stent was completed with the event confirmed as resolved.Per the investigator the cause of the gutter endoleak was possibly related to a previous intervention/index procedure.No additional clinical sequelae were reported and the patient is fine.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO ANCHOR SYSTEM - HELI-FX AAA
Common Device Name
ENDOVASCULAR SUTURING SYSTEM
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
MDR Report Key8877659
Report Number9612164-2019-03298
Device Sequence Number1948562
Product Code OTD
UDI-Device Identifier00643169647695
UDI-Public00643169647695
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K140036
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,study
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/27/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2018
Device Model NumberSA-85
Device Catalogue NumberSA-85
Device Lot Number50039015
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/12/2019
Supplement Date Received by Manufacturer09/27/2019
Initial Report FDA Received Date08/09/2019
Supplement Report FDA Received Date09/27/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age72 YR
Patient Weight87
-
-