• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH NO 3. TRIATHLON TS PLUS TIBIAL INSERT X3 POLY 9MM; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH NO 3. TRIATHLON TS PLUS TIBIAL INSERT X3 POLY 9MM; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Catalog Number 5537-G-309
Medical Device Problem Codes Improper or Incorrect Procedure or Method (2017); Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Injury (2348)
Date of Event 06/07/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.The information in this report was provided by stryker orthopaedics legal affairs department.No additional information is available at this time due to the ongoing litigation.Should additional information become available, the evaluation summary will be submitted in a supplemental report.The following devices were also listed in this report: triathlon asymmetric x3 patella; cat# 5551-g-320; lot# ymay.Triathlon ps fem component, cemented; cat# 5515-f-301; lot# nfrwa.Tri ts baseplate size 3; cat# 5521-b-300; lot# tvwhb.It cannot be determined which, if any of these devices may have caused or contributed to the patient's experience.Device not returned.
 
Event or Problem Description
It was reported through the communication of an attorney that allegedly the patient was revised due to failed left total knee arthroplasty.
 
Additional Manufacturer Narrative
An event regarding incorrect selection involving a triathlon patella was reported.The event was confirmed via medical review.Method & results: product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.Medical records received and evaluation: a review of the provided medical information by a clinical consultant indicated: multiple procedure-related factors including poor femoral cementation technique with malalignment of the knee in excessive valgus further contributing to an overload condition and the pain magnified by patellofemoral overstuffing have in concert contributed to immediate fixation failure of the triathlon ps/ts femoral component within the first few weeks post primary surgery on (b)(6) 2016, causing severe and persistent pain with functional limitations and secondary instability ultimately requiring revision surgery with exchange of all triathlon devices to depuy revision knee devices on (b)(6) 2018.No device-related factors are associated with any of the implanted devices at any time.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: the medical review revealed that the patella was found too thick with 28-mm, which was determined to be a procedural related factor which contributed to immediate fixation failure of the triathlon ps/ts femoral component.No device-related factors are associated with any of the implanted devices at any time.No further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
It was reported through the communication of an attorney that allegedly the patient was revised due to failed left total knee arthroplasty.Update on 26-nov-2019 by qs based on medical review: left knee revision surgery was performed on (b)(6) 2018, under the diagnosis of ¿failed left knee arthroplasty¿ with exchange of all stryker components for depuy (non stryker) revision knee devices.At exploration, the instability was confirmed and the patella was found too thick with 28-mm.The femoral component was deemed loose as it could be ¿popped out¿ without effort.Only 7-weeks post surgery, the second x-rays of (b)(6) 2016, confirmed already significant proximal migration of the femoral component with more than 1-cm subsidence compared with the previous x-ray of (b)(6) 2016, with lucency and bone resorption along almost the entire femoral implant-bone interface.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NO 3. TRIATHLON TS PLUS TIBIAL INSERT X3 POLY 9MM
Common Device Name
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key8878689
Report Number0002249697-2019-02779
Device Sequence Number18892717
Product Code MBH
UDI-Device Identifier07613327027129
UDI-Public07613327027129
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K141056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2018
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/23/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/30/2017
Device Catalogue Number5537-G-309
Device Lot NumberMLP80H
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/19/2019
Initial Date Received by Manufacturer 07/15/2019
Supplement Date Received by Manufacturer11/26/2019
Initial Report FDA Received Date08/09/2019
Supplement Report FDA Received Date12/23/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age54 YR
-
-