| Model Number |
25AGFN-756 |
| Medical Device Problem Code |
Difficult to Insert (1316)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/16/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Further information regarding this event has been requested.The results/method and conclusion codes along with investigation results will be provided in a subsequent submission.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2019, a 25 mm regent heart valve was selected for implant.During implant, the physician was unable to get the valve to rotate and exchanged the valve for another unknown valve.Due to the valve unable to rotate, there was a clinically significant delay in procedure.No patient consequences were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of inability to rotate the valve could not be confirmed.The investigation confirmed the device met functional specifications when analyzed at abbott.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications at the time of commercialization.The cause of the reported event could not be conclusively determined.
|
| |
|