| Model Number |
27MJ-501 |
| Medical Device Problem Code |
Gas/Air Leak (2946)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/15/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An event of the valve not completely opening could not be confirmed.Morphological and hydrodynamic examination indicated the valve met abbott specifications.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications at the time of commercialization.Hydrodynamic testing at the time of manufacturing and upon return to abbott indicated the valve functioned normally.This test ensures proper leaflet coaptation and hemodynamic performance.The cause of the reported event could not be conclusively determined.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2019, a 27 mm valve was selected for implant.During implant procedure the physician noted that the valve could not completely open.The valve was exchanged for a 27 mm valve from medtronic and a clinically significant delay in procedure time was reported.No patient consequences reported.
|
| |
|