RESPIRONICS, INC. EVERFLO OXYGEN CONCENTATOR; GENERATOR, OXYGEN, PORTABLE Back to Search Results
Model Number 1020000
Device Problems Mechanical Problem (1384); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Low Oxygen Saturation (2477)
Event Date 07/24/2019
Event Type  Injury  
Event Description
The manufacturer received information alleging an everflo oxygen concentrator was emitting an odor and was not working correctly.The patient was admitted to the hospital.The investigation is still ongoing.A follow up report will be submitted when the manufacturer has completed the investigation.
 
Manufacturer Narrative
The manufacturer previously reported an everflo oxygen concentrator that allegedly was emitting an odor and was not working correctly.The patient was admitted to the hospital.The device was returned to the manufacturer's service center for evaluation.The device ws tested and was found to deliver oxygen to design specifications.There was no evidence of a burning odor being emitted by the device and there was no evidence of thermal damage to the device.The manufacturer concludes the device did not cause or contribute to the patient's hospitalization.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVERFLO OXYGEN CONCENTATOR
Type of Device
GENERATOR, OXYGEN, PORTABLE
Manufacturer (Section D)
RESPIRONICS, INC.
175 chastian meadows court
kennesaw GA 30144 3724
MDR Report Key8879612
MDR Text Key153851565
Report Number1040777-2019-00038
Device Sequence Number1
Product Code CAW
Combination Product (y/n)N
PMA/PMN Number
K061261
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 08/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/09/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number1020000
Device Catalogue Number1020000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/13/2019
Date Manufacturer Received09/13/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
-
-