• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD ACCESS SYSTEMS POWERLOC SAFETY INFUSION SET 20G X 0.75IN WITH Y-SITE; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS POWERLOC SAFETY INFUSION SET 20G X 0.75IN WITH Y-SITE; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number N/A
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 07/17/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has not been returned to the manufacturer for evaluation.A lot history review (lhr) of ascys0071 showed no other similar product complaint(s) from this lot number.
 
Event or Problem Description
It was reported that when opening the kit, it was noticed that the cover of needle was not contained in it.There was no reported patient involvement.
 
Additional Manufacturer Narrative
The following were reviewed as part of this investigation: patient severity, complaint and lot history review, applicable previous investigation(s), sample (if available), labeling, and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of a missing component is inconclusive due to the state of the returned sample.One 20 g x 0.75 in.Powerloc winged infusion set with y-site in an opened package was returned for evaluation.An initial visual observation showed obvious evidence of use.The needle shaft was observed to be slightly bent at its base.The needle guard was not returned with the sample.The opened state of the packaging made it difficult to assess when and where the component went missing.Possible causes include improper kit packaging and misplacement of kit components after opening.A lot history review (lhr) of ascys0071 showed no other similar product complaint(s) from this lot number.
 
Event or Problem Description
It was reported that when opening the kit, it was noticed that the cover of needle was not contained in it.There was no reported patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERLOC SAFETY INFUSION SET 20G X 0.75IN WITH Y-SITE
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key8879676
Report Number3006260740-2019-02361
Device Sequence Number11336498
Product Code FPA
UDI-Device Identifier00801741151477
UDI-Public(01)00801741151477
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K060812
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 09/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number0672034J
Device Lot NumberASCYS0071
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/16/2019
Event Location Hospital
Initial Date Received by Manufacturer 07/18/2019
Supplement Date Received by Manufacturer08/26/2019
Initial Report FDA Received Date08/09/2019
Supplement Report FDA Received Date09/11/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-