• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DAVOL INC., SUB. C.R. BARD, INC. -1213643 VENTRALIGHT ST MESH; SURGICAL MESH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVOL INC., SUB. C.R. BARD, INC. -1213643 VENTRALIGHT ST MESH; SURGICAL MESH Back to Search Results
Catalog Number 5954600
Medical Device Problem Codes Defective Device (2588); Material Deformation (2976); Insufficient Device Problem Information (3190); Migration (4003)
Health Effect - Clinical Codes Adhesion(s) (1695); Hernia (2240); Injury (2348)
Date of Event 04/19/2016
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
At this time no conclusions can be made.The patient's attorney alleges injuries and subsequent surgery; however, no details have been provided.No lot number has been provided; therefore a review of the manufacturing records is not possible.Should additional information be provided a supplemental emdr will be submitted.Not returned.
 
Event or Problem Description
Attorney alleges the patient underwent surgery for implant of an unspecified bard/davol ventralight st on (b)(6) 2016.It is alleged that the patient had subsequent surgical intervention due to the hernia mesh device(s).As reported, the patient is making a claim for an adverse patient outcome against the bard/davol ventralight st.As reported, the attorney alleges patient experienced emotional distress and the device was defective.
 
Additional Manufacturer Narrative
At this time no conclusions can be made.The patient's attorney alleges injuries and subsequent surgery; however, no details have been provided.No lot number has been provided; therefore a review of the manufacturing records is not possible.Addendum: this supplemental emdr is submitted to document additional information provided.Root cause is inconclusive; no conclusion can be made as to the extent to which the implant may have caused or contributed to the patient's postoperative course.Note, the manufacturing date (15-sep-2015) is considered to be a best estimate.Note: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
Attorney alleges the patient underwent surgery for implant of an unspecified bard/davol ventralight st on (b)(6) 2016.It is alleged that the patient had subsequent surgical intervention due to the hernia mesh device(s).As reported, the patient is making a claim for an adverse patient outcome against the bard/davol ventralight st.As reported, the attorney alleges patient experienced emotional distress and the device was defective.Addendum per additional information provided: (b)(6) 2016 ¿ patient was diagnosed with incarcerated ventral incisional hernia thereby underwent robotic assisted repair with the implant of ventralight st (device #1).Per op notes, ¿a piece of round ventralight st mesh (device #1) was placed intraperitoneally and sutured.¿ (b)(6) 2016 ¿ patient was diagnosed with recurrent incarcerated incisional hernia thereby underwent laparoscopic repair with the implant of ventralight st (device #2).Per op notes, ¿adhesiolysis was done.The upper portion of mesh (device #1) had pulled through and folded over.There was a small portion of the falciform ligament was incarcerated in the defect.A piece of ventralight st mesh (device #2) was in right upper quadrant and overlapped with other piece of mesh (device #1).¿.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VENTRALIGHT ST MESH
Common Device Name
SURGICAL MESH
Manufacturer (Section D)
DAVOL INC., SUB. C.R. BARD, INC. -1213643
100 crossings blvd.
warwick RI 02886
Manufacturer (Section G)
BARD SHANNON LIMITED 3005636544
san geronimo industrial park
lot #1, road #3, km 79.7
humacao PR 00791
Manufacturer Contact
andrew topoulos
100 crossings blvd.
warwick, RI 02886
4018258495
MDR Report Key8879691
Report Number1213643-2019-06870
Device Sequence Number254408
Product Code FTL
UDI-Device Identifier00801741031632
UDI-Public(01)00801741031632(17)170928(10)HUZI0387
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K101851
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 06/24/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/28/2017
Device Catalogue Number5954600
Device Lot NumberHUZI0387
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/02/2019
Supplement Date Received by Manufacturer06/11/2025
Initial Report FDA Received Date08/09/2019
Supplement Report FDA Received Date06/24/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/15/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age61 YR
Patient SexFemale
-
-