| Catalog Number |
XT16JS65NSC006S |
| Medical Device Problem Codes |
Break (1069); Fracture (1260)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Information (3190)
|
| Date of Event |
05/30/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Information was received indicating that the left eyelet to a smiths medical portex bivona flextend¿ tts¿ tracheostomy tube was observed to break during routing trach site care.It was reported that the reported to be "sticky and gummy following 5 sterilization cycles.The tube was reported to be used a total of six times and sterilized five times for the past year.Subsequently, the trach tube was immediately changed out with no further adverse effects.
|
| |
|
Event or Problem Description
|
|
Information was received that incident occurred during routine trach care, which includes cleaning of the tracheostomy site.A tracheostomy tube change out occurred immediately.
|
| |
|