• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: W.L. GORE & ASSOCIATES GORE VIABAHN ENDOPROSTHESIS - 3; NIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W.L. GORE & ASSOCIATES GORE VIABAHN ENDOPROSTHESIS - 3; NIP Back to Search Results
Catalog Number VBJR051502A
Medical Device Problem Code Difficult or Delayed Activation (2577)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/17/2019
Type of Reportable Event Malfunction
Event or Problem Description
The following information was reported to gore: on (b)(6) 2019 a patient was undergoing treatment of a superficial femoral artery in-stent stenosis.The intent was to reline the stent with a gore® viabahn® endoprosthesis.When the device was advanced to the target location, deployment was attempted, but after the device was deployed half way, the deployment line broke.The delivery catheter was then cut to try to expose the remaining deployment line.The delivery catheter was peeled open and the deployment line was exposed.The deployment line was pulled, and deployment was completed.Upon removal of the delivery system it was noted the deployment line was knotted.A second device was deployed to complete the treatment and the patient outcome was good.
 
Additional Manufacturer Narrative
H.6.Results code 2: 213: the engineering evaluation stated the following: part of the deployment line attached to the deployment knob, and two sections of the delivery catheters were returned.The deployment line appeared to be broken.It measured 149cm with a 7.5cm single fiber coming from the end.There were two knots in the deployment line approximately 50.5cm and 55cm from the deployment knob.The delivery catheter appeared to be cut and resulted in 2 pieces.The pieces measured 32cm and 93cm.Approximately 39cm of the section of delivery catheter that included the distal tip was cut down the middle.Engineering evaluation conclusion(s) are: deployment line broken.Based on the device examination performed, no manufacturing anomalies were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE VIABAHN ENDOPROSTHESIS - 3
Common Device Name
NIP
Manufacturer (Section D)
W.L. GORE & ASSOCIATES
flagstaff AZ
MDR Report Key8880306
Report Number2017233-2019-00607
Device Sequence Number10819751
Product Code PFV
UDI-Device Identifier00733132623921
UDI-Public00733132623921
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
P040037
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/22/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/02/2020
Device Catalogue NumberVBJR051502A
Device Lot Number17459604
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/15/2019
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date08/09/2019
Supplement Report FDA Received Date09/13/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-