| Model Number |
400CE |
| Medical Device Problem Code |
Insufficient Heating (1287)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/10/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
User reported the hemotherm 400ce would not heat over 29c.Complaint #(b)(4)was opened and is being investigated.No adverse event was reported.
|
| |
|
Event or Problem Description
|
|
User reported the hemotherm 400ce heater-cooler was heating slowly.The unit would not heat over 29c.No adverse event was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated and determined to have a defective heater.The heater was replaced to correct.
|
| |
|