| Model Number |
MMT-523RNAH |
| Medical Device Problem Codes |
Connection Problem (2900); Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/14/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Customer reported via phone call that the insulin pump had cosmetic damage and crack around battery compartment.Customer's blood glucose level was unknown.Customer reported that reservoir was not able to lock in place.Customer was advised to revert to backup plan.Insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device received unable to perform the displacement due to complete broken reservoir tube lip, cracked battery tube threads and cracked reservoir tube window noted.The p-cap not locks into the reservoir compartment properly due to completely broken reservoir tube lip noted.
|
| |
|