• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. REF THREE HOLE SHELL 52MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. REF THREE HOLE SHELL 52MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS Back to Search Results
Catalog Number 71336452
Medical Device Problem Code Connection Problem (2900)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/26/2019
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the liner cannot match with the acetabular cup.Backup device was used.No injury reported.Delay greater than 30 min reported.
 
Additional Manufacturer Narrative
It was reported that the liner cannot match with the acetabular cup.The associated reflection 3 hole shell was returned and evaluated.A dimensional evaluation was completed on the returned device for all critical-to-quality features, and the failure mode could not be confirmed through this evaluation.No obvious signs of damage was visually observed.Our investigation including a review of the manufacturing records for the listed batch did not reveal any deviation from the standard manufacturing processes.A review of the complaint history for the listed part revealed no prior complaints for the listed failure mode with the same batch number.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A relationship, if any, between the device and the reported incident could not be corroborated.No further investigation warranted for this complaint; however smith and nephew will continue to monitor for future complaints and investigate as necessary.Should additional information be received, the complaint will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REF THREE HOLE SHELL 52MM
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key8883002
Report Number1020279-2019-03010
Device Sequence Number2716069
Product Code MBL
UDI-Device Identifier03596010454355
UDI-Public03596010454355
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K960094
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/07/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number71336452
Device Lot Number18GM05117
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 07/26/2019
Supplement Date Received by Manufacturer11/01/2019
Initial Report FDA Received Date08/12/2019
Supplement Report FDA Received Date11/07/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-