• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. BRAUN DOMINICIAN REPULIC INC. BLOOD ADMINISTRATION SET; SET, ADMINISTRATION, INTRA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN DOMINICIAN REPULIC INC. BLOOD ADMINISTRATION SET; SET, ADMINISTRATION, INTRA Back to Search Results
Catalog Number NF5140
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 07/24/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).Although it was confirmed that the device involved is not available for evaluation, the investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
 
Event or Problem Description
As reported by the user facility: event 1: customer reported two different issues with the blood tubing leaking from one side or the other.No injury reported.
 
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).Although no sample was provided for further evaluation, all available information was forwarded to our supplier for further evaluation.Based on the supplier's statement, the supplier reviewed their manufacturing process and were able to observe inconsistencies during the assemble process.It was determined that the issue of leaking is due to insufficient glue when adhering the molded ends to the tube.The supplier is currently generating new visual pictures and written instructions to aid during the assembly process.All operators involved with the assembly process will be trained to these new procedures.We will maintain this report for future reference, and continue to monitor other reports for similar occurrences.A review of the discrepancy management system (dsms) database was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.In addition, review of the batch history records was also performed for the reported lot number and no non-conformances or deviations were noted during the manufacturing process or final inspections.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BLOOD ADMINISTRATION SET
Common Device Name
SET, ADMINISTRATION, INTRA
Manufacturer (Section D)
B. BRAUN DOMINICIAN REPULIC INC.
las americas industrial park
km22 autopista las americas
DR 
MDR Report Key8883044
Report Number9614279-2019-00117
Device Sequence Number11353234
Product Code FPA
UDI-Device Identifier04046964187483
UDI-Public04046964187483
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/22/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2024
Device Catalogue NumberNF5140
Device Lot Number0061681745
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/24/2019
Supplement Date Received by Manufacturer07/24/2019
Initial Report FDA Received Date08/12/2019
Supplement Report FDA Received Date09/22/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-