• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC COOK SILICONE BALLOON HYSTEROSALPINGOGRAPHY INJECTION CATHETER; HES INSUFFLATOR-CARBON DIOXIDE, URETOTUBAL (AND ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC COOK SILICONE BALLOON HYSTEROSALPINGOGRAPHY INJECTION CATHETER; HES INSUFFLATOR-CARBON DIOXIDE, URETOTUBAL (AND ACCESSORIES) Back to Search Results
Model Number G17870
Medical Device Problem Code Deflation Problem (1149)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A follow up report will be submitted should additional relevant information become available.
 
Event or Problem Description
It was reported that after a hysterosalpingography was completed, the health care provider attempted to remove the cook silicone balloon hysterosalpingography injection catheter that was used during the procedure.There were issues removing the catheter from the uterus due to the balloon not deflating.They were eventually able to deflate the balloon.Initial reporter reported this event occurred on four different occasions.(patient reference identifier: (b)(6)).No adverse event was reported due to this malfunction.
 
Event or Problem Description
No additional event or patient information is available at this time.
 
Additional Manufacturer Narrative
Investigation evaluation: reviews of complaint history, device history record, manufacturing instructions, quality control data, and the instructions for use (ifu) were conducted during the investigation.The complaint device was not returned; therefore, no physical examinations could be performed.The complaint device was not returned; therefore, no physical examinations could be performed; however, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.A review of the device history record found no non-conformances related to the reported failure mode.A review of complaint history records shows three other complaints associated with the complaint device lot.The other three cases are related to the same failure mode, but due to the individual nature of the manufacturing and inspection processes for the devices in this lot, it is unlikely that these events are an indication of an issue with the entire lot.The instructions for use (ifu), provides the following information to the user related to the reported failure mode: how supplied: upon removal from the package, inspect the product to ensure no damage has occurred.A review of relevant manufacturing documents was conducted.It was concluded that the device aspect in question was visually/functionally inspected by quality control and no related gaps in production or processing controls were noted.The complainant did not return the complaint device for investigation.At the time of the complaint investigation, all products in this lot have been distributed to customers, so no representative product was available from the lot for evaluation.The complaint was confirmed based on the customer¿s testimony.The cause of the issue is unknown.Per the quality engineering risk assessment, no further action is warranted.Cook medical has notified the appropriate personnel and will continue to monitor this device via the complaints database for similar complaints.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COOK SILICONE BALLOON HYSTEROSALPINGOGRAPHY INJECTION CATHETER
Common Device Name
HES INSUFFLATOR-CARBON DIOXIDE, URETOTUBAL (AND ACCESSORIES)
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key8883733
Report Number1820334-2019-01985
Device Sequence Number13960173
Product Code HES
UDI-Device Identifier00827002178700
UDI-Public(01)00827002178700(17)220109(10)9432727
Combination Product (Y/N)N
PMA/510(K) Number
K891290
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,other,use
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/09/2022
Device Model NumberG17870
Device Catalogue NumberJ-CHSG-703000
Device Lot Number9432727
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/01/2019
Supplement Date Received by Manufacturer08/28/2019
Initial Report FDA Received Date08/12/2019
Supplement Report FDA Received Date09/05/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-