| Catalog Number |
328510 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/29/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Date of event: unknown.The date received by manufacturer has been used for this field.Medical device expiration date: unknown.Device manufacture date: unknown.Investigation summary: level a investigation complaint evaluation / complaint history check for the event(s) that occurred.Severity: s_1__; occurrence: unable to perform complaint lot history check for plunger rod difficult to move due to unknown lot number.Investigation summary: no samples (including photos) were returned therefore the complaint could not be confirmed and the root cause is undetermined.If samples are received in the future the complaint will be reopened for further investigation.Unable to perform dhr review due to unknown lot number.Investigation conclusion: as no samples and/or photo(s) were received the investigation concluded: unconfirmed: bd was not able to duplicate or confirm the customer¿s indicated failure as no samples or photos were returned.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.Root cause description: root cause cannot be determined at this time as the issue is unconfirmed as no samples or photos were returned.Rationale: based on the investigation, no additional investigation and no capa is required at this time.
|
| |
|
Event or Problem Description
|
|
It was reported that relion® insulin syringe plunger rod is difficult to move.This was discovered during use.The following information was provided by the initial reporter: material no.: 328510, batch no.: unknown.It was reported by the consumer that the plunger rod is difficult to move.Verbatim: consumer reported plunger rod difficult to move before injection.Problem occurred a few months ago.
|
| |
|
Search Alerts/Recalls
|