• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INNOVA VASCULAR; STENT, SUPERFICIAL FEMORAL ARTERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INNOVA VASCULAR; STENT, SUPERFICIAL FEMORAL ARTERY Back to Search Results
Model Number 26926
Medical Device Problem Codes Activation, Positioning or Separation Problem (2906); Material Deformation (2976); Positioning Problem (3009)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/22/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Age at time of event: (b)(6).
 
Event or Problem Description
It was reported that the stent jumped forward and elongated.The 90% stenosed target lesion was located in the highly calcified and normally tortuous left common iliac and external iliac arteries.An ipsilateral approach was used to access the lesion.A bsc 6f super sheath was used for access and a non-bsc wire was selected.Pre-dilation was performed with a 5mm balloon, however, pre-dilation did not open the lesion large enough.A 7 x 150 x 130 innova vascular stent was selected for use within the left common iliac to the distal portion of the external iliac artery.During deployment, the stent jumped forward and elongated.When beginning to deploy, the stent moved forward through a high grade stenosis of the left common iliac.The stent deployed without issue for 30-40mm, but then once the stent hit the high grade stenosis it elongated for about 30mm before returning back to looking normal and not elongated.No stent fracture was noted.Due to the elongation, the end of the stent completed deployment within the sheath.By pulling the sheath back a few millimeters, the stent was able to deploy and properly expand inside of the common femoral artery.The procedure was completed by performing a planned post dilation of the stent.No patient complications were reported.
 
Event or Problem Description
It was reported that the stent jumped forward and elongated.The 90% stenosed target lesion was located in the highly calcified and normally tortuous left common iliac and external iliac arteries.An ipsilateral approach was used to access the lesion.A bsc 6f super sheath was used for access and a non-bsc wire was selected.Pre-dilation was performed with a 5mm balloon, however, pre-dilation did not open the lesion large enough.A 7 x 150 x 130 innova vascular stent was selected for use within the left common iliac to the distal portion of the external iliac artery.During deployment, the stent jumped forward and elongated.When beginning to deploy, the stent moved forward through a high grade stenosis of the left common iliac.The stent deployed without issue for 30-40mm, but then once the stent hit the high grade stenosis it elongated for about 30mm before returning back to looking normal and not elongated.No stent facture was noted.Due to the elongation, the end of the stent completed deployment within the sheath.By pulling the sheath back a few millimeters, the stent was able to deploy and properly expand inside of the common femoral artery.The procedure was completed by performing a planned post dilation of the stent.No patient complications were reported.
 
Additional Manufacturer Narrative
A2: age at time of event: 78 years old device evaluated by mfr: the complaint device was not received at the complaint investigation site for analysis.A photo provided by the customer showed what appeared to be two implanted stents, but no measurements were provided, so the length of the stent could not be confirmed.The left stent appeared to be under-deployed in the bottom section.Under-deployed stents often are due to the struts hitting a hard surface or calcium.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INNOVA VASCULAR
Common Device Name
STENT, SUPERFICIAL FEMORAL ARTERY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key8884016
Report Number2134265-2019-09538
Device Sequence Number11442080
Product Code NIP
UDI-Device Identifier08714729874133
UDI-Public08714729874133
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
P140028
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/10/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/21/2022
Device Model Number26926
Device Catalogue Number26926
Device Lot Number0023517671
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/22/2019
Supplement Date Received by Manufacturer08/19/2019
Initial Report FDA Received Date08/12/2019
Supplement Report FDA Received Date09/10/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GUIDEWIRE: STIFF GLIDEWIRE BY TERUMO; GUIDEWIRE: STIFF GLIDEWIRE BY TERUMO; GUIDEWIRE: STIFF GLIDEWIRE BY TERUMO
-
-