The associated device was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The weld, just below the handle on the device, fractured causing the device to disassemble and become inoperable.All pieces were returned.The handle was damaged, just above the fracture site, with numerous nicks from impacts.The device was manufactured in 2017 and exhibits signs of extensive wear/usage.This failure was likely due to fatigue loading of the weld joint caused by repeated impacts.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.
|