It was reported that there was a low air leak and low flow on the arctic sun device.
The patient temperature was 38c with a target temperature of 36.
5c.
The flow rate was 1.
1l/min, and the water temperature was 5.
6c but the nurse stated the pads did not feel cold on the patient.
Per troubleshooting with ms&s, the nurse stated that the patient only had two universal pads on because the unit did not have a full set of pads available.
The nurse was advised to locate a full set of pads to properly deliver therapy.
The nurse stated she would check in another unit or neighboring hospitals for a full set of pads.
Per follow up with nurse (b)(6) via phone on 22jul2019, the patient was continuing therapy successfully with no further issues with a new set of pads.
|