| Model Number |
8900-0223-01 |
| Medical Device Problem Code |
Failure of Device to Self-Test (2937)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during biomed testing, the associated defibrillator failed self test using these electrode pads.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The electrodes were returned for evaluation; the malfunction was duplicated during visual evaluation.The malfunction was attributed to a disconnected jumper wire (self test wire) on the electrode pad.This type of failure does not disable the electrode from delivering therapy when attached to a defibrillator.This is a malfunction which only impacts self-test of the electrode with the defibrillator.This type of malfunction does not pose any clinical impact and does not meet our guidelines for reportability.Analysis for reports of this type has not identified an increase in trend.
|
| |
|
Search Alerts/Recalls
|