| Model Number |
ABI541 |
| Medical Device Problem Code |
Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
05/20/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device was explanted due to a suspected device failure.Analysis found a hermetic seal failure.The conclusion from device analysis is that the loss of function was caused by moisture penetrating the hermetic seal and impacting the device function.This report is submitted on august 13, 2019.
|
| |
|
Event or Problem Description
|
|
Per the clinic, the patient was explanted on (b)(6) 2019 because of a suspected device failure.The patient was re-implanted with a device at the same surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: the correct date of awareness (b4) and (g4) should be 06 june 2019, not 18 july 2019 as previously reported.This report is submitted on 6 march 2020.
|
| |
|