ANIMAS LLC ANIMAS VIBE; INSULIN INFUSION PUMP Back to Search Results
Device Problem Retraction Problem (1536)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
 
Event Description
On (b)(6) 2019, the reporter contacted animas, alleging a motor (rewind issue) issue.It was reported that the rewind stops prior to completion.There was no indication that the product caused or contributed to an adverse event.This complaint is being reported because the issue may result in an inability to use the product which may lead to long term cessation of delivery.
 
Manufacturer Narrative
Device evaluation: the pump has been returned and evaluated by product analysis on 30-oct-2019 with the following findings: a review of the pump¿s black box showed cs 078 motor rewind alarm.During testing, a cs 078 alarm occurred during the rewind step.The pump was opened and an internal moisture found on the motor flex connector and surrounding components.A crack in the battery compartment was observed.Leak testing revealed a leak at the battery compartment.Animas has conducted a review of the device history record for this pump and confirmed that it was operating within required specifications at the time of release.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANIMAS VIBE
Type of Device
INSULIN INFUSION PUMP
Manufacturer (Section D)
ANIMAS LLC
965 chesterbrook blvd
wayne PA 19087
Manufacturer (Section G)
ANIMAS LLC
965 chesterbrook blvd
wayne PA 19087
Manufacturer Contact
kristen lopolito
965 chesterbrook blvd
wayne, PA 19087
MDR Report Key8886236
MDR Text Key154099180
Report Number2531779-2019-04351
Device Sequence Number1
Product Code OYC
Combination Product (y/n)Y
Reporter Country CodeCA
PMA/PMN Number
P130007
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,foreign
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 08/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/12/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/29/2019
Was the Report Sent to FDA? Yes
Date Manufacturer Received08/03/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age36 YR
-
-