Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, the device had cuff inflation and deflation issue.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: one sample of device was received for evaluation.The reported condition was confirmed.An inflation/deflation test was performed and it was observed the cuff deflated immediately.A visual inspection was conducted and it was noticed the cuff presents a cut which was measured 0.9915 inches.Information has been added to the database and trends will continue to be monitored.If information is provided in the future, a supplemental report will be issued.
|
| |
|