| Catalog Number |
636006P |
| Medical Device Problem Codes |
Use of Device Problem (1670); Device Handling Problem (3265)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/16/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No known impact or consequence to patient: no reported significant delay to procedure, patient consequence or harm.Blushing/bleeding was repaired with application of bioglue and patient reported as stable post op.Information from the customer indicated that the device was not rinsed in saline for 5 minutes prior to use.Device handling problem - ifu states that failure to rinse for 5 minutes could lead to the graft being more susceptible to leakage when implanted.Vascutek ltd, now considers this complaint closed.Further action is not planned, however, the issue will be tracked and trended as part of the on-going complaints trending and reporting process and if an adverse trend develops action may be taken at that time.
|
| |
|
Event or Problem Description
|
|
The event was reported as follows: after suturing of the graft and graft supply restoring, there was leakage through the graft noticed.Doctor waited for 15 minutes but the oozing did not stop.The glue was applied to stop leakage.
|
| |
|