| Model Number |
MMT-722NAS |
| Medical Device Problem Codes |
No Display/Image (1183); Device Difficult to Program or Calibrate (1496); Moisture or Humidity Problem (2986)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/04/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the insulin pump had stuck button alert and insulin pump was not working.Customer¿s blood glucose was 70 mg/dl.Customer stated that buttons not responding.Customer stated that insulin pump was exposure to moisture.Troubleshooting was performed.Customer was advised that the insulin pump needed to be replaced and to discontinue use of the insulin pump and revert to a back-up plan.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device received with blank display and constant tone due to moisture damage on the electronics assembly.Unable to perform all functional testing including the operating currents, a21 error test, rewind, basic occlusion, occlusion, prime or a33 and excessive no delivery test or verify button keypad anomaly due to blank display.Device received with moisture damage motor, vibrator and battery tube assembly.
|
| |
|