• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NEODENT - JJGC S.A. TI TITAMAX IMPLANT (5.0)4.5X9; ENDOSSEOUS DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NEODENT - JJGC S.A. TI TITAMAX IMPLANT (5.0)4.5X9; ENDOSSEOUS DENTAL IMPLANT Back to Search Results
Catalog Number 109.327
Medical Device Problem Code Loss of Osseointegration (2408)
Health Effect - Clinical Code Tissue Damage (2104)
Date of Event 06/11/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Considering the surgical methodology, it was seen that the dentist has selected an implant design which is not recommended for the patient's bone quality.Moreover, it was seen that the dentist has used more drills than recommended to perform the implant installation.The investigation of the complaint was carried out considering the analysis of the information given by the dentist in the guarantee form, visual conference of the product returned by the dentist and the evaluation of relevant production records.
 
Event or Problem Description
(b)(4).The dentist reported that 1 year and 2 months after the dental implant was installed in intraoral region 20#, its loss of osseointegration was observed.Moreover, the patient presented insufficient bone quality/quantity (bone type iii).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TI TITAMAX IMPLANT (5.0)4.5X9
Common Device Name
ENDOSSEOUS DENTAL IMPLANT
Manufacturer (Section D)
NEODENT - JJGC S.A.
av. juscelino kubitschek de
oliveira, 3291 cic
curitiba PR 81270 -200
BR  81270-200
MDR Report Key8887475
Report Number3008261720-2019-04022
Device Sequence Number11376847
Product Code DZE
UDI-Device Identifier07898237567129
UDI-Public07898237567129
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K101207
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2019
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 08/13/2019
Report Date (Section F) 07/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/27/2022
Device Catalogue Number109.327
Device Lot Number800300153
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/19/2019
Was the Report Sent to FDA? No
Distributor Facility Aware Date08/13/2019
Device Age26 MO
Event Location Other
Date Report to Manufacturer07/19/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/19/2019
Initial Report FDA Received Date08/13/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age56 YR
-
-