| Catalog Number |
101214010 |
| Medical Device Problem Code |
Degraded (1153)
|
| Health Effect - Clinical Codes |
Inflammation (1932); Pain (1994); Loss of Range of Motion (2032); No Code Available (3191)
|
| Date of Event |
11/13/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Initial reporter occupation: lawyer.(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Patient was revised to address pain.Additional information received via ppf and medical records.In additional to what were previously alleged, ppf alleges pseudotumor, clunking sound, inflammation, dislocation and decreased range of motion.After review of medical records,patient was revised to address adverse reaction to metal debris.Operative notes indicated small spot of corrosion at the trunnion with no obvious trunnionosis.Stem added to the complaint for the alleged corrosion.Added medical history and patient's age.Doi: (b)(6) 2008; dor: (b)(6) 2017, left hip.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary = > no device associated with this report was received for examination.A worldwide complaint database search found no other related reported incidents against the provided product code/lot number combination since release for distribution.Based on previous investigations, this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|