| Model Number |
1205328 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Blood Loss (2597)
|
| Date of Event |
07/13/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The customer indicated that the device is available to return for investigation; it has not been received.
|
| |
|
Event or Problem Description
|
|
The event involved a report stating that the extension tubing has disconnected from the microbore t connector part; it has separated from inside the connector.It was not pulled on, not damaged, it just suddenly disconnected and was attached to a central line.The baby started bleeding out of the t- connector.The event was noticed when the nurse did the rounds in nicu.There was patient involvement, and "some blood loss but nothing permanent".
|
| |
|
Additional Manufacturer Narrative
|
|
The complaint of separation 120530480 could not be confirmed by investigation.No product samples, pictures, or videos were received for investigation.Without the return of the sample a comprehensive failure investigation cannot be performed and a cause cannot be determined.A batch record review was performed to lot# 69194ns, list# 12053-0480 and no discrepancies that may have contributed to a complaint of this nature were found.The quality inspection of final visual and physical evaluation results indicated that the product met specification requirements.
|
| |
|