|
The ht70 ventilator was returned to medtronic for evaluation.An investigation was performed and the circuit check failures were con firmed in the event history logs only.The popping noise issue could not be duplicated.A different issue was observed during testing, where the unit did not pass the system leak test.It was observed that the safety valve was partially pressed outside the manifold body.When the safety valve was pressed back into the manifold body and the pump/manifold assembly was re-installed, the unit ran with no alarms or errors.When the manifold assembly was replaced with a known working part, the unit continued to not pass the system leak test (although the leak value was lower).The pump assembly was removed and examined.No tears were observed in the diaphragm.A small amount of contamination/debris was observed under the check valves.The o-ring on the outlet end of the pump was observed to be damaged.The system leak was determined to potentially be caused by the safety valve pressed out of the manifold body, contamination under the check valves of the pump assembly, and a damaged o-ring.The pump/manifold assembly was noted to require replacement and the unit was going to be processed under service instructions.The unit would be returned to the loaner pool once the repairs are completed.If information is provided in the future, a supplemental report will be issued.
|