• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HOLOGIC, INC AFFIRM BREAST BIOPSY GUIDANCE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOLOGIC, INC AFFIRM BREAST BIOPSY GUIDANCE SYSTEM Back to Search Results
Model Number STLC-00004
Device Problem Imprecision (1307)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/24/2019
Event Type  malfunction  
Event Description
It was reported that during two biopsy exams the needle went through the patient breast and hit the detector.No medical intervention required.It was determined that the needle data needed to be entered after a software update.An applications specialist was able to assist the customer to enter the needle data.Once this was completed the system is working as intended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFFIRM BREAST BIOPSY GUIDANCE SYSTEM
Type of Device
BIOPSY GUIDANCE SYSTEM
Manufacturer (Section D)
HOLOGIC, INC
36 & 37 apple ridge road
danbury CT 06810
Manufacturer Contact
kristin fornieri
36 & 37 apple ridge road
danbury, CT 06810
2037318491
MDR Report Key8888555
MDR Text Key154145782
Report Number1220984-2019-00090
Device Sequence Number1
Product Code IZH
UDI-Device Identifier15420045505858
UDI-Public15420045505858
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K122836
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 07/24/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/13/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSTLC-00004
Device Catalogue NumberSTLC-00004
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/24/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/01/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-