| Model Number |
366997 |
| Medical Device Problem Code |
Failure to Advance (2524)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/23/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A pantera leo balloon was chosen to treat a severe calcified lesion (83 percent stenosis degree) in a severe tortuous lad.Despite repeated attempts it was not possible to cross the lesion.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned instrument was subjected to a technical analysis and the corresponding production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.The technical investigation of the returned instrument showed that the balloon has not been inflated but slightly unfolded.According to the provided information the remaining lumen diameter in the target lesion was smaller than the crossing profile of the balloon.Review of the production documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.Based on the conducted investigations of the device being subject to this complaint, no material or manufacturing related root cause could be determined.The final root cause is most likely related to the patients anatomy.
|
| |
|