COVIDIEN 5.0FR URETHANE UMB CATH; CATHETER, UMBILICAL ARTERY Back to Search Results
Model Number 8888160341
Medical Device Problem Codes Fluid/Blood Leak (1250); Migration (4003)
Health Effect - Clinical Code No Information (3190)
Date of Event 07/12/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device history record review could not be performed because a lot number was not received with the complaint.As part of our manufacturing process, all device history records are reviewed and approved by quality, prior to release of product.Samples were not received for the investigation.Because a sample was not returned, we were unable to perform a follow up investigation to include functional and visual evaluations to confirm the defect and root cause analysis.If a sample is returned; this complaint will be reopened.A corrective action is not applicable at this time.Functional testing and visual inspections are being performed according to our current quality standards and inspection procedures.This complaint will be used for qa tracking and trending purposes.
 
Event or Problem Description
The customer stated that the catheter is too soft and bends on insertion and moves out of place (migrates).The customer further stated that during one incident the catheter leaked into the abdomen.No additional incident details were able to be provided by the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5.0FR URETHANE UMB CATH
Common Device Name
CATHETER, UMBILICAL ARTERY
Manufacturer (Section D)
COVIDIEN
edificio b20 calle #2 zona fra
alajuela
CS 
MDR Report Key8890771
Report Number3009211636-2019-00666
Device Sequence Number3039204
Product Code FOS
UDI-Device Identifier20884527005120
UDI-Public20884527005120
Combination Product (Y/N)N
Initial Reporter CountrySF
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 08/13/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model Number8888160341
Device Catalogue Number8888160341
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/15/2019
Initial Report FDA Received Date08/13/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-