(b)(4).A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Investigation summary: it was reported that the device had a breakage needle.One empty opened box and twenty-two unopened samples of product code vcp322, lot jd7526 were returned for analysis.The complaint sample was not received for evaluation.During the visual inspection of thirteen samples, no defects were found on the packages.The samples were opened, and the swage and attachment area were noted to be as expected; also, the needles were examined under magnification and no defects, bending needle or needle breakage were found.The sutures were dispensed without problem and examined along of strand, no defects or damaged were observed.Besides, the samples were tested by resistance test to needle and met the requirements.The manufacturing records were reviewed, and the manufacturing/packaging criteria were met prior to the release of this batch.
|