| Model Number |
ZXT150 |
| Medical Device Problem Code |
Unintended Movement (3026)
|
| Health Effect - Clinical Codes |
Halo (2227); No Code Available (3191)
|
| Date of Event |
07/09/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The intraocular lens (iol) is not returning for evaluation; therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, and complaint trending for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.(b)(4).All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a zxt150 16.0 diopter intraocular lens (iol) was implanted in the patient's right (od) eye on (b)(6) 2019.At 1 day postop, the lens rotated > 10 degree resulted in unexpected refraction.Consequently, patient complained of halos and glare.The lens was explanted on (b)(6) 2019 and replaced with the same model at a lower diopter, 15.0.Patient has recovered post-operatively.The customer also indicated that the lens is not available for return.No further information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
|
| |
|