| Model Number |
5866-38M |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); High impedance (1291); Impedance Problem (2950)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/26/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Concomitant medical products: 5071-35 lead; 5071-35 lead, implanted: (b)(6) 2017.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that both the left ventricular (lv) lead and adapter had high impedance measurements.It was noted that the lv tip electrode impedance had a high degree of variability with isometric left arm movements.It was also noted that noise was present on the electrogram (egm).It was noted that reprogramming was done, and both leads were turned off.No patient complications have been reported as a result of this event.
|
| |
|