MAUDE Adverse Event Report: CAREFUSION, INC BD TRAY EPID CONT WE18G3.5 C20 LOR5 S/L/L-E; ANESTHESIA CONDUCTION KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION, INC BD TRAY EPID CONT WE18G3.5 C20 LOR5 S/L/L-E; ANESTHESIA CONDUCTION KIT Back to Search Results
Catalog Number 406047
Medical Device Problem Code Device Markings/Labelling Problem (2911)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 07/25/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used for this field.Medical device expiration date: unknown.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.Device manufacture date: unknown.
 
Event or Problem Description
It was reported that the incorrect needle length was in the tray with a bd tray epid cont we18g3.5 c20 lor5 s/l/l-e.This occurred on 2 separate occasions prior to use.The following information was provided by the initial reporter: it was reported that two trays contained the incorrect needle length.Health care professional called to report that doctor found 2 kits that had 19 g instead of 20g lot is unknown.
 
Additional Manufacturer Narrative
Investigation: no samples were returned for analysis.The root cause could not be confirmed without a lot number or sample but the investigation noted that packaging of the 406047 is a manual process.As part of this manufacturing process, all components are to be verified with the applicable bill of material at start up and at any time during the manufacturing process when more components are delivered to the manufacturing line.The investigation identified the most probable root cause to be a failure in the component verification step with the applicable bill of materials during the manufacture of the affected finished good lot.Based on the complaint investigation, the most probable root cause was that manufacturing personnel did not properly verify components as they were delivered to the manufacturing line.Based on the complaint investigation, the most probable root cause was that manufacturing personnel did not properly verify components as they were delivered to the manufacturing line.
 
Event or Problem Description
It was reported that the incorrect needle length was in the tray with a bd tray epid cont we18g3.5 c20 lor5 s/l/l-e.This occurred on 2 separate occasions prior to use.The following information was provided by the initial reporter: it was reported that two trays contained the incorrect needle length.Health care professional called to report that doctor found 2 kits that had 19 g instead of 20g lot is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD TRAY EPID CONT WE18G3.5 C20 LOR5 S/L/L-E
Common Device Name
ANESTHESIA CONDUCTION KIT
Manufacturer (Section D)
CAREFUSION, INC
400 east foster rd
mannford OK 74044
MDR Report Key8895032
Report Number1625685-2019-00090
Device Sequence Number1658719
Product Code CAZ
UDI-Device Identifier00382904060473
UDI-Public00382904060473
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
DISCRETION
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,other,use
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number406047
Device Lot NumberUNKNOWN
Initial Date Received by Manufacturer 07/25/2019
Supplement Date Received by Manufacturer07/25/2019
Initial Report FDA Received Date08/14/2019
Supplement Report FDA Received Date09/05/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-